The Titamed UNI-Smile Distractor® - Surgical Procedure

Click here to download the surgical procedure as pdf in ENGLISH (paper A4 EN)
Click here to download the surgical procedure as pdf in SPANISH (paper A4 ES)
Click here to download the surgical procedure as pdf in GERMAN (paper A4 DE)
Click here to download the surgical procedure as pdf in FRENCH (paper A4 FR)
Click here to download the surgical procedure as pdf in DUTCH (paper A4 NL)

Placement UNI-Smile Distractor: Live Surgery

*The sequence of surgery

1. Corticotomies: lateral - midline + posterior disjunction
2. Placement of the UNI-Smile Distractor with self-drilling screws between teeth n° (14)-15-16 and (24)-25-26

 

Incision

 

A cross palatal incision or an L-incision, between premolar and molar or between premolars has shown to be the ideal location. The incision is made, 1 cm downwards as indicated on the picture.

Placement: Blue mark ALWAYS to the left!
 

The abutment plates are placed horizontally. The blue side is intended for the left side of the patient. The Smile Distractor must be fixed to the palatal bone with a self-drilling screw on each side (DSSA06 or DSSA07 - screwdriver IT.0315).

Activation
 

The surgeon must activate the Smile Distractor until the two front teeth have an opening of ± 1mm. To do this, place the key instrument (IT.0314) over the middle part of the distractor body and turn from the front to the back

Blocking screw
 

The blocking screw (D.002) is placed in one of the two available threaded holes with the screwdriver (IT.0128). Place the blocking screw perpendicular over the cylinder, with a clear view of the screw hole. The blocking screw should be fixed real firm, using wrist force.

*Tip

Place a tampon during placement in case the blocking screw should drop in the mouth.

 

*THE SEQUENCE OF THE DISTRACTION

1. At the end of the surgery the blocking screw is tightly engaged, using wrist force. Place a tampon in case the blocking screw should drop.
2. One week after the surgery the activation of the Smile Distractor is started. Remove the blocking screw and expand untill bone resistance is felt, then turn once or twice more. During activation (1 to 2 weeks) the patient should turn two times a day a quarter of a turn. At each quarter of a turn a new color appears.
3. After completion of the distraction put the blocking screw back real firm, using wrist force. The distractor needs to be in place 3 to 6 months.