The Titamed UNI-Smile Distractor EXTRA® - Surgical Procedure
Click here to download the surgical procedure as pdf in ENGLISH (paper A4 EN) ![]()
Click here to download the surgical procedure as pdf in DUTCH (paper A4 NL) ![]()
|
||
Incision |
||
|
The most appropriate position for the incision is between the premolar and molar or between the premolars. A cross palatal incision can be made. The correct length of the distractor can be measured from bone to bone with the assistance of the dummies. | |
Placement: Blockingnut (blue) always on the left!
|
||
|
Place the Smile Distractor plates horizontally. The side of the Blocking nut is meant for the left side. Fix the Smile Distractor with a self-drilling screw between two roots on each side. (DSSA06 or DSSA07 - screwdriver IT.0315). | |
Activation |
||
|
The surgeon activates the Smile Distractor so that an opening is created of ± 0.5 to 1mm between the two central incisors. This is done by placing the key instrument (IT.0414) over the mid-section of the distractor and turning from front to back. |
|
Blockingnut |
||
![]() |
Turn the locking nut against the module of the distractor using the key instrument (IT.0414). It is important that the blocking nut does not remain in the gums during the surgery. | |
|
||